High-Resolution Transducers up to 1,000 MHz

With a frequency of 1,000 MHz, our transducers offer the world’s highest applied frequency for acoustic mapping and analysis. This means that resolutions down to 0,8 µm µm can be achieved on suitable samples. PVA TePla Analytical Systems GmbH manufactures transducers of 100–1,000 MHz for the most diverse high-resolution applications, based on sapphire lens technology and ZnO piezo materials of the active element.

These special lenses were developed in particular for the analysis of thin layers, but also for quantitative acoustic microscopy. In this case, Rayleigh waves are induced just below the surface of the material and analyzed using v(z) and f(v) curves. By means of suitable Fourier analysis, surface wave velocities (Rayleigh waves) can be measured with an accuracy exceeding 0.1%, which enables local determination of the modulus of elasticity or shear modulus.

Product Information

Find in-depth information about the possibilities of ultrasound inspection, our systems, transducers, and software as well as specific application possibilities.

[Translate to English:] PVA TePla Analytical Systems Transducer Development

Generation of the electrical matching network to ensure optimum adaptation to the necessary application.

[Translate to English:] PVA TePla Analytical Systems Transducer

PVA TePla Analytical Systems Transducer

PVA TePla Analytical Systems Ultrasonic System Transducer

Transducer / Range

The table shows a list of available transducers. Additional versions are available on request and can be manufactured for specific applications.

No. Frequency [MHz] Axial resolution [µm] Detection limit [µm] Focal length in water [mm] Additionally available focal lengths (in water) [mm] Aperture diameter [mm]
1 Transducer 5 MHz 257.18 85.73 37.00 20 mm – 60 mm (customer-specific), or non-focused 3.5–6.36
2 Transducer 10 MHz 160.74 53.58 19.00 20–50 mm (customer-specific), or non-focused 3.5–6.36
3 Transducer 15 MHz 98.91 32.97 37.00 20–50 mm (customer-specific), or non-focused 3.5–6.36
4 Transducer 20 MHz 75.64 25.21 10.00 20–50 mm (customer-specific), or non-focused 2.5–5
5 Transducer 25 MHz 58.45 19.48 25.00 25 mm (customer-specific), or non-focused 2.5–5
6 Transducer 30 MHz 49.46 16.49 13.00 12.7 mm 4
7 Transducer 35 MHz 42.86 14.29 30.00 30 mm 4
8 Transducer 35 MHz 42.86 14.29 35.00 40 mm 4
9 Transducer 50 MHz 29.90 9.97 12.00 10–12 mm 3.0–4.0
10 Transducer 75 MHz 20.09 6.70 12.70 8–12.7 mm 3
11 Transducer 80 MHz 19.78 6.59 10.00 8–12 mm 3
12 Transducer 100 MHz 14.95 4.98 8.00 8 mm or 10 mm 2
13 Transducer 110 MHz 13.68 4.56 8.00 8 mm 2
14 Transducer 125 MHz 11.91 3.97 9.00 7–9 mm 2
15 Transducer 150 MHz 9.97 3.32 5.90 5.9 mm 1.85–2.2
16 Transducer 150 MHz 9.97 3.32 8.00 8 mm 1.85–2.2
17 Transducer 150 MHz 9.97 3.32 4.00 4 mm 1.85–2.2
18 Transducer 150 MHz 9.97 3.32 3.20 3.2 mm 1.85–2.2
19 Transducer 150 MHz 9.97 3.32 8.0 8 mm 1.85–2.2
20 Transducer 175 MHz 8.57 2.86 4.00 4 mm 1.85–2.2
21 Transducer 175 MHz 8.57 2.86 3.20 3.2 mm 1.85–2.2
22 Transducer 175 MHz 8.57 2.86 8.00 8 mm 1.85–2.2
23 Transducer 200 MHz 7.52 2.51 8.00 8 mm 1.85–2.2
24 Transducer 200 MHz 7.52 2.51 4.00 4 mm 1.85–2.2
25 Transducer 200 MHz 7.52 2.51 8.00 8 mm 1.85–2.2
26 Transducer 200–250 MHz 6.01 2.00 5.00 5 mm 1.85–2.2
27 Transducer 200–250 MHz 6.01 2.00 4.00 4 mm 1.85–2.2
28 Transducer 200–250 MHz 6.01 2.00 3.20 3.2 mm 1.85–2.2
29 Transducer 200–250 MHz 6.01 2.00 8.00 8 mm 1.85–2.2
30 Transducer 300 MHz 5.02 1.67 3.20 3.2 mm 1.85–2.2
31 Rayleigh wave lens 100 MHz 7.56 2.52 1.80 1.8 mm 1.8
32 Rayleigh wave lens 200 MHz 3.76 1.25 0.60 0.6 mm 0.6
32 Rayleigh wave lens 400 MHz 1.88 0.63 0.30 0.3 mm 0.3
33 Rayleigh wave lens 1,000 MHz 0.60 2.52 0.08 0.08 mm 0.08
34 Rayleigh wave lens 800–1,300 MHz 0.60 1.25 0.06 0.06 mm 0.06
35 Rayleigh wave lens 1,300–2,000 MHz 0.60 0.63 0.06 0.06 mm 0.06

Immediate Contact

slideoutcontact
Please complete all fields marked with *.
Please solve the following question:
captcha

PVA TePla Analytical Systems GmbH
Deutschordenstrasse 38
73463 Westhausen, Germany

Phone: +49 7363 9544 0
Fax: +49 7363 9544 113

Request Product Information

Please complete all fields and choose the product you would like to learn more about. The product data sheet will then be e-mailed to you. Please understand that in order to preserve our natural resources, we do not send out any printed material.

Product Information_info
Please complete all fields marked with *.

I am aware that when I make contact using this contact form, my data will be stored to facilitate the further processing of my inquiry and subsequent inquiries.

Please solve the following question:

Please rate our Service

Daten
Please rate our service operation / service technician according to the following criteria:
(1 star deficient, 5 stars very good)
Please solve the following question:
captcha

Please rate our training

Daten
Please rate our training according to the following criteria:
(0 points deficient, 5 points very good)
Please solve the following question:
captcha